Pathological Insights on Polypropylene Mesh Complications From Laparoscopic  Sacrocolpopexy: A Case Series

Okui, N. & Okui, M.A.
Cureus, 16(3), p.e56354., 2024.

Background:

The use of polypropylene mesh in laparoscopic sacrocolpopexy (LSC) is a common treatment for pelvic organ prolapse (POP). Despite its widespread application, postoperative complications such as mesh pain and infection sometimes necessitate the removal of the mesh. However, it remains unclear in which cases mesh removal is warranted. Our research focused on the pathological changes at the sacral fixation point of the mesh. We sought to evaluate the pathological alterations of the sacral mesh removed through an innovative approach of transvaginal natural orifice transluminal endoscopic surgery (vNOTES).

Materials and Methods:

This retrospective study included nine patients who underwent mesh removal surgery at the Yokosuka Urogynecology and Urology Clinic in 2023. Extraction surgery was performed using vNOTES with the GelPoint Access Platform (Applied Medical JAPAN HEADQUARTERS, Tokyo, Japan). Non-ablative Erbium YAG and Neodymium YAG lasers (RenovaLase, SP Dynamis; Fotona d.o.o., Ljubljana, Slovenia) were utilized for persistent stress urinary incontinence, fecal incontinence, vaginal erosion, and bleeding after surgery. Patients were categorized based on mesh fixation conditions, including unintended mesh overlap (Group I), excessive traction (Group II), and signs of mesh aging (Group III). This categorization helped to understand the distinct pathological outcomes associated with each condition.

Results:

Pathological findings from the mesh removed via vNOTES varied significantly across the groups. In Group I, characteristic large vacuole formation and accumulation of atypical giant cells were observed, attributed to mesh overlap. Group II presented with vacuole formation, fiber degradation, and tissue destruction as a result of excessive mesh traction. In Group III, the aging of the mesh was marked by cracks in the surrounding tissues and granuloma formation. These detailed observations provide crucial insights into the underlying causes of mesh-related pain and other complications, highlighting the complexity of bodily responses to mesh implants.

Conclusion:

This study demonstrated the effectiveness of vNOTES for polypropylene mesh removal in patients with post-LSC complications, resulting in significant pain reduction. Pathological analysis revealed that mesh-related issues stem from the surgical techniques, mesh properties, and long-term bodily reactions. These findings provide valuable insights for improving mesh design and POP treatment strategies. Despite the technical challenges, vNOTES is recommended for mesh removal in patients with pain. Additionally, the combination of UEL, VEL, targeted laser irradiation, AEL, and Nd:YAG laser treatments showed promising results in managing post-mesh removal complications such as stress urinary incontinence, vaginal erosion, bleeding, and fecal incontinence, offering hope for improved patient outcomes.

Keywords: polypropylene mesh, mesh removal, laparoscopic sacrocolpopexy, pelvic organ prolapse, transvaginal natural orifice transluminal endoscopic surgery

LA&HA Projects